| A | B | C | D | E | F | G | H | I | J | |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | Safe Lancet by Smith's Medical | Acti Lance | Contact Activated Lancet by BD | |||||||
| 2 | Yes | No | No answer | Yes | No | No answer | Yes | No | No answer | |
| 3 | Question 2 | 82% | 18% | 96% | 4% | 83% | 17% | |||
| 4 | Question 3 | 82% | 12% | 6% | 92% | 8% | 100% | |||
| 5 | Question 4 | 71% | 23% | 6% | 88% | 8% | 4% | 100% | ||
| 6 | Question 5 | 36% | 58% | 6% | 83% | 17% | 94% | 6% | ||
| 7 | Question 6 | 100% | 88% | 12% | 100% | |||||
| 8 | Question 7 | 41% | 47% | 12% | 70% | 26% | 4% | 94% | 6% | |
| 9 | Question 8 | 59% | 41% | 42% | 58% | 83% | 17% | |||
| 10 | ||||||||||
| 11 | Question 2 | The product/equipment is easy to use | ||||||||
| 12 | Question 3 | The product/equipment is safe to use in a pediatric setting | ||||||||
| 13 | Question 4 | The product/equipment is clinically and developmentally appropriate for use in infants | ||||||||
| 14 | Question 5 | The product/equipment performs its intended purpose and is acceptable | ||||||||
| 15 | Question 6 | Adequate training was provided in order to use this product/equipment | ||||||||
| 16 | Question 7 | Patient satisfaction/comfort will improve or remain the same by using this product/equipment | ||||||||
| 17 | Question 8 | I would recommend this product/equipment for use | ||||||||
| 18 | ||||||||||
| 19 | Color codes | |||||||||
| 20 | highest approval rating | |||||||||
| 21 | middle approval rating | |||||||||
| 22 | lowest approval rating | |||||||||
| A | B | C | D | E | |
|---|---|---|---|---|---|
| 1 | Needle device evaluation form: January 09 | ||||
| 2 | Product or Equipment: Safe Lancet by Smiths Medical | ||||
| 3 | Project Leader: Myra Horn RN, BSN, MHSA, MBA | ||||
| 4 | Number of responses: 17 | ||||
| 5 | Times device was used: < 5 times = 11 5-10 times = 5 >10 times = 1 | ||||
| 6 | |||||
| 7 | Yes | No | no answer | Additional comments | |
| 8 | I was appropriately trained on this device | 53% | 47% | ||
| 9 | The product/equipment is easy to use | 82% | 18% | awkward, hard to push | |
| 10 | The product/equipment is safe to use in a pediatric setting | 82% | 12% | 6% | pressure may allow for lancet to go deeper than intended |
| 11 | The product/equipment is clinically and developmentally appropriate for use in infants | 71% | 23% | 6% | possible to go too deep |
| 12 | The product/equipment performs its intended purpose and is acceptable | 36% | 58% | 6% | Bleeds too slowly, no blood was obtained, poor flow |
| 13 | Adequate training was provided in order to use this product/equipment | 100% | |||
| 14 | Patient satisfaction/comfort will improve or remain the same by using this product/equipment | 41% | 47% | 12% | squeezing on fingers to bleed, pt had to be stuck twice |
| 15 | I would recommend this product/equipment for use | 59% | 41% | finger does not bleed well, does not perform well | |
| 16 | Major advantages of this product/equipment: | It is safety | |||
| 17 | Good blow flow | ||||
| 18 | Good for safety | ||||
| 19 | It is easier to grasp and you have more control of the device | ||||
| 20 | Easy to use, patient bleeds enough for collection | ||||
| 21 | Great blood return | ||||
| 22 | None | ||||
| 23 | 1.8mm great on adult fingers | ||||
| 24 | Have to press | ||||
| 25 | Major disadvantages of this product/equipment: | Poor flow | |||
| 26 | Not a good sample | ||||
| 27 | did not penetrate enough to bleed properly | ||||
| 28 | didn't provide adequate capillary bleed on children | ||||
| 29 | Bllod does not flow freely, finger slow to bleed, have to force blood out | ||||
| 30 | awkward | ||||
| 31 | Hard to activate poor flow | ||||
| 32 | None | ||||
| 33 | Major disadvantages of this product/equipment cont: | Didn't bleed well | |||
| 34 | Not deep enough leading to multiople sticks to obtain specimen | ||||
| 35 | too much pressure required to release, causing inccreased pressure on site | ||||
| 36 | Degree of pressure may cause deeper penetration in finger | ||||
| 37 | Would you recommend that we try another product? | Yes, could cause problems down the road | |||
| 38 | Yes, haven't found a better device yet | ||||
| 39 | Yes, not a good blood flow | ||||
| 40 | Yes | ||||
| 41 | Yes, product does not perform effectively | ||||
| 42 | No I think this is the product that we all like | ||||
| 43 | No | ||||
| 44 | yes not appropraite for pediatrics | ||||
| 45 | Comments | No good!!!! | |||
| 46 | Very good | ||||
| 47 | I wish we could keep using this lancet in primary care | ||||
| 48 | Love this one!! | ||||
| 49 | Recommend only the purple one if necessary | ||||
| 50 | the children like this product because they did not feel the stick | ||||
| 51 | As a phlebotomist I did not like this product because I could not get a good stick | ||||
| 52 | Not a good device to work with | ||||
| 53 | |||||
| A | B | C | D | E | |
|---|---|---|---|---|---|
| 1 | Needle device evaluation form: January 09 | ||||
| 2 | Product or Equipment: Acti Lance | ||||
| 3 | Project Leader: Myra Horn RN, BSN, MHSA, MBA | ||||
| 4 | Number of responses: 19 | ||||
| 5 | Times device was used: < 5 times = 13 5-10 times = 5 >10 times = 1 | ||||
| 6 | |||||
| 7 | Yes | No | no answer | Additional comments | |
| 8 | I was appropriately trained on this device | 42% | 58% | ||
| 9 | The product/equipment is easy to use | 96% | 4% | Had to press down too hard | |
| 10 | The product/equipment is safe to use in a pediatric setting | 92% | 8% | twisting feature is a choking hazard, small pieces | |
| 11 | The product/equipment is clinically and developmentally appropriate for use in infants | 88% | 8% | 4% | not as effective |
| 12 | The product/equipment performs its intended purpose and is acceptable | 83% | 17% | did not get adequate blood flow, did not perform well | |
| 13 | Adequate training was provided in order to use this product/equipment | 88% | 12% | did not get adequate blood flow, did not perform well | |
| 14 | Patient satisfaction/comfort will improve or remain the same by using this product/equipment | 70% | 26% | 4% | did not perform on every use, left bruise, pt didn't like |
| 15 | 42% | 58% | did not perform on every use, painful left bruise | ||
| 16 | multiple use to collect sample, product hurt | ||||
| 17 | I would recommend this product/equipment for use | pt finger does not bleed well, did not provide adequate coll. | |||
| 18 | Major advantages of this product/equipment: | none | |||
| 19 | Easy to use | ||||
| 20 | Works well | ||||
| 21 | Easy to hold on to and loose | ||||
| 22 | Patient fells little to know pain | ||||
| 23 | Easy to hold on to and loose | ||||
| 24 | The 17 gauge would be a more versatile lancet, It could be used or heel sticks | ||||
| 25 | or finger sticks of older children | ||||
| 26 | Works well with finger sticks | ||||
| 27 | When did perform, it worked very well on blood flow | ||||
| 28 | None | ||||
| 29 | Good for finger sticks | ||||
| 30 | Blood flowed well from foot for me | ||||
| 31 | No improvement over current product, more painful than BD or Saf T lance | ||||
| 32 | Easy to use, pt didn't complain | ||||
| 33 | Major disadvantages of this product/equipment: | Not comfortable for phlebotomist or patient | |||
| 34 | Had to press too hard | ||||
| 35 | Blood does not flow freely | ||||
| 36 | Depth of cut, bulky | ||||
| 37 | Did not perform on every use | ||||
| 38 | Painful and left bruise at finger puncture site | ||||
| 39 | hard to control | ||||
| 40 | Can't place it on the feet so it can cut the foot | ||||
| 41 | When compared to other trial lancets that were 21 gauge or less, blood was | ||||
| 42 | obtained | ||||
| 43 | Would you recommend that we try another product? | Yes continue with Tenderlett | |||
| 44 | no | ||||
| 45 | No, Acti Lance | ||||
| 46 | Phlebotomist needs blood to flow better | ||||
| 47 | Yes just so that we can compare | ||||
| 48 | This product is difficult to keep still and puncture skin | ||||
| 49 | No I like these | ||||
| 50 | Yes staff not satisfied | ||||
| 51 | Would also like to suggest we look for a product without | ||||
| 52 | the twist off piece which is a choking hazard | ||||
| 53 | Yes, should not be painful | ||||
| 54 | Comments | There were 3 different needle depths and I was not informed for the one to use | |||
| 55 | for the age of the child | ||||
| 56 | Most of my blood sample clot using this product. Blood does not flow as fast | ||||
| 57 | as needed. I did not have problem with other equipment | ||||
| 58 | Used all 3 sizes on 4 patients none of them even said ouch, got good blow flow | ||||
| 59 | with all three | ||||
| 60 | Used only 17 gauge | ||||
| 61 | Used 17 and 23 gauge | ||||
| 62 | Makes noise on release that will cause patient to jerk - normal response | ||||
| 63 | Could cause a fear rather than puncture, could result in a jagged slice of finger | ||||
| 64 | |||||
| A | B | C | D | E | |
|---|---|---|---|---|---|
| 1 | Product or Equipment: BUZZY | ||||
| 2 | Project Leader: Myra Horn RN, BSN, MHSA, MBA | ||||
| 3 | Number of responses: 9 | ||||
| 4 | Times device was used: < 5 times = 5 5-10 times =1 >10 times = 3 | ||||
| 5 | |||||
| 6 | Yes | No | no answer | Additional comments | |
| 7 | I was appropriately trained on this device | 100% | |||
| 8 | The product/equipment is easy to use | 100% | |||
| 9 | The product/equipment is safe to use in a pediatric setting | 89% | |||
| 10 | The product/equipment is clinically and developmentally appropriate for use in infants | 89% | Not for use with very small infants | ||
| 11 | The product/equipment performs its intended purpose and is acceptable | 89% | not effective | ||
| 12 | Adequate training was provided in order to use this product/equipment | 100% | |||
| 13 | Patient satisfaction/comfort will improve or remain the same by using this product/equipment | 100% | |||
| 14 | I would recommend this product/equipment for use | 89% | |||
| 15 | |||||
| 16 | Major advantages of this product/equipment: | Pain control | |||
| 17 | Reduces pain for needle sticks and allows easier draws of blood | ||||
| 18 | Provides distraction to children age 3-10 some of the 2 year olds were afraid | ||||
| 19 | of the BEE. Does help with the pain less patients jerked their arm as | ||||
| 20 | needle was inserted so much have reduced pain as that is normal response | ||||
| 21 | No expiration date and easy to use | ||||
| 22 | Keeps the patients mind off the needle stick | ||||
| 23 | It's a distraction technique. The kids seem to like the bee | ||||
| 24 | Distracts pediatric patient during procedure | ||||
| 25 | Major disadvantages of this product/equipment: | Vibration | |||
| 26 | Parental stress, sometimes parents are sure the won't work before you try | ||||
| 27 | Makes the child uncomfortable and cold .. Makes the tips numb | ||||
| 28 | Can't be used on infants | ||||
| 29 | Sometimes it is not appropriate if things need to be done in a hurry | ||||
| 30 | Some of the kids did not like the cold pack | ||||
| 31 | Some pediatric patients did not care for buzzing on the arm | ||||
| 32 | No real disadvantage, but I did think that the frozen packs were too cold | ||||
| 33 | for younger children. I tried getting them to "thaw slightly" or using | ||||
| 34 | a paper towel between the cold pack and the arm, this seemed | ||||
| 35 | to be better received and did not leave red spot on the arm from the cold | ||||
| 36 | Would you recommend to try another product | No, | |||
| 37 | Comments | The lab was able use Buzzy more than the nurses. Although the | |||
| 38 | med techs were slow to start using the device, once they got used to it | ||||
| 39 | they used it more and more. The nurses did not want to take the time | ||||
| 40 | to use it. Not effective on leg intramuscular shots. | ||||
| 41 | It makes older children uncomfortable, cold | ||||
| 42 | Really think this is a bonus for the patient, would like the flippet cards to be | ||||
| 43 | larger so they would present visual of the needle stick. I found that if I | ||||
| 44 | asked the questions for the flippet cards it works much better, the parents | ||||
| 45 | don't seem to be able to multi task with holding the child and reading the | ||||
| 46 | questions. | ||||
| 47 | |||||